Maintenance and Care
Cleaning the Cylinders and the Sampling Chamber
4.2 Cleaning the Cylinders and the Sampling Chamber
Clean the surface of the CASY Instrument regularly using a damp microfiber cloth. It is important to
remove salt solution spillages in the instrument immediately. To do this, raise the sample platform
and remove it by pulling forwards. Use 80% ethanol to remove organic impurities. Under no circum-
stances, however, should other organic solvents be used.
4.3 How to take the CASY Instrument Out of Operation
If your CASY System will not to be operated for several weeks, the fluid system should be drained.
Before shutting down, the fluid system must also be cleaned (see Chapter 4.1,
Weekly Cleaning Rou-
tine
). Please always follow this procedure to shut down the system:
1
Carry out steps
1
to
10
of the
Weekly cleaning routine
.
2
Replace the two CASY cups containing distilled water with two empty CASY cups.
3
Select the
Service Menu
option from the main menu.
4
Perform the
Dry Liquid System
command from the
Service
Menu
.
5
Empty the waste container and the two CASY cups.
6
You can now switch off your system. To make sure that the instrument is completely shut down,
disconnect the power cable from the power inlet.
7
Ensure that the system is adequately protected from dust for the duration it is shut down.
8
You should begin the
Start-up
procedure with the
Weekly cleaning routine
.
You must ensure that when the fluid system of the CASY Instrument contains a saline solution, it
is not allowed to dry out. This is very important to prevent malfunction and damage to the fluid
system and valves caused by salt crystals.