94MS36 Rev. F2 S/W Ver 2.06
13-25
increases to regulate the lamp. At idle, this voltage is typically 45 - 90V~
and depends on the lamp age and ambient temperature. When an exposure
at high power is initiated, the lamp voltage rises to the maximum. The
voltage remains there after the lamp returns to low power, until the lamp
temperature again stabilizes.
•
#6
The shutter voltage is present during the rotation of the shutter.
The shutter motor runs on 120V~ measured to terminal 4.
•
#7
Shutter position DC common (see terminal 9)
•
#8
Not presently used
•
#9
The shutter position switch reports the position of the shutter to
the printed circuit board. This switch is open when the shutter is closed,
and closes when the shutter opens. This signal is low voltage DC and
can be measured (+) on terminal 9, (-) on terminal 7. The meter will read
+12 VDC when the shutter is closed, and 0 VDC when the shutter is
opened.
•
#10
Interlock and thermostat switching. During normal operation,
terminal 10 is shorted to terminal 4 (V~ common). An V~ voltage present
would indicate the glass switch or one of the two thermostat switches
open.
•
#11
Not presently used.
•
#12#14
Incoming voltage. To get load voltage, meter between
terminals 12 & 14 while power is on. On 208/240V~ equipment set
voltage selection switch accordingly, (low=219V~ or below, high=220V~
or above).
•
#13
a. Autoformer
Light Function States
Model
Warm-Up
Idle
Low Exp.
Med Exp.
High Exp.
Cool Down
Trigger
Unit Fans
On / On
On / On
On / On
On / On
On / On
On / On
On / On
Power Relay
On / On
On / On
On / On
On / On
On / On
Off / Off
Off / Off
Shutter
Closed / Off
Closed / Off
Open / Off
Open / Off
Open / Off
Open / Off
* / Off
Lamp Fans
Off / Off
Var.Slow / On Slow/Med. / On Med./Fast / On
Fast / On
Fast / On
Off / Off
High Relay
On / On
Off / Off
Off / Off
Off / Off
On / On
* / *
* / *
Medium Relay
On / On
Off / Off
Off / Off
On / On
On / On
* / *
Warm-Up
- Occurs after the lamp ignites and ends when the lamp has
warmed to operating temperatures. The lamp fans are held off and the
unit goes to high power. Due to the technology of the power supply, the
units do not draw large amounts of current during warm-up. Due to the
restrictive cooling, the lamps warm up rapidly, without surge current.
The shutter will not open until the unit reaches temperature.
Содержание Accuprint AP30-5000
Страница 1: ...P P AP30 5000 AP30 8000 High Resolution Printer Service Manual 94MS36 Rev F2 S W Ver 2 06 O ...
Страница 70: ...11 4 94MS36 Rev F2 S W Ver 2 06 ...
Страница 72: ...12 2 94MS36 Rev F2 S W Ver 2 06 ...
Страница 136: ...16 10 94MS36 Rev F2 S W Ver 2 06 ...
Страница 138: ...O Corporation 17112 Armstrong Avenue Irvine Ca U S A 92614 U S A 949 399 6500 Fax 949 399 6501 ...