42
Heater Temperature is no longer displayed, as illustrated in Figure 28 Moreover, in the
Objective Heater sub Menu the User can change the offset and select the option ‘Reset to
factory’ to reset Objective Heater Settings to the default values. (See Figure 29)
Set Point 37.0°C
Base
37.0
HM 37.0
Lid
37.0
Figure 28. Homepage when objective Heater is disabled.
Menu 1/2
Select an Adapter
Offset
à
Objective Heater
8
Objective Heater 1/2
à
Enabled: OFF
Offset:
0.0
Ramp °C/min 0.5
Objective Heater 2/2
à
Reset to factory
8
Click on Reset To
Factory
Reset to factory ready to be
changed with flashing Item
Confirm or
Exit
8
-
â
Figure 29. Reset to factory default
In this Submenu, the User can set a manual offset for Objective Heater as described in Figure
30 and following the procedure described in paragraph 14.
Menu 1/2
Select an Adapter
Offset
à
Objective Heater
8
Objective Heater 1/2
Enabled:
OFF
à
Offset:
0.0
Ramp °C/min 0.5
8
Enter to change the
Offset Value
Confirm or
Exit
8
-
â
á
Up/Down until to
desired Offset Value
Figure 30. How to change the Objective Heater offset.
From the main menu press the ‘Menu’ key on the front panel.
Move the cursor to ‘Objective Heater’ using the Arrow Keys.
Press ‘Enter’