Alarm Settings Submenu
8
Alarm Settings 1/2
à
Alarm:
On
Buzzer:
Off
Alarm Dev,°C: 1.0
Alarm Settings 2/2
à
Dev Time,min: 04
Reset to factory
Factory Default: N
Alarm Settings 2/2
Dev Time,min: ##
à
Reset to factory
Factory Default: N
8
Select the value
to be changed
Move the cursor to the
Parameter to be
changesd
â
á
8
Confirm
new value
Change
the value
â
á
-
Exit
8
Click on Reset To
Factory
Reset to factory ready to be
changed with flashing Item
Confirm or
Exit
8
-
Menu 2/2
à
Alarm Settings
Screen Settings
Product info
Figure 34. Alarm Settings Menu.
An ‘Alarm’ can be enabled (‘on’) or disabled (‘off’). With the ‘Alarm’ ‘on’ a ‘Buzzer’ can also be
enabled. Right after turning the device on or right after changing the set point temperature,
the ‘Alarm’ will be in standby for 30 minutes, allowing the incubator to reach stable conditions.
After 30 minutes the ‘Alarm’ will activate and start operating. The Alarm function will check
for alarm conditions. An exclamation mark (!) will appear next to the component of the
chamber with temperature different from the set point more than the Alarm Dev and for a
length of time longer than Dev. Time. (See Figure 35).
Set Point 37.0°C
Base!
34.0
HM 37.0
Lid
37.0
OBH 37.0
Figure 35. Visual alarm Active.
If the ‘Buzzer’ is enabled the sound alarm will go off as well. When ‘Alarm’ is active and
‘Buzzer’ is on, an audible alarm is active. Press the ‘Up’ key one time to stop the buzzer alarm
only, for 30 seconds only. If the alarm condition persists, after 30 seconds, the buzzer will go
off again.