Changing Set Points
.
à
Set Point 25.0°C
Base 25.0 HM 25.0
Lid 25.0 OBH 25.0
8
à
Set Point
37.0°C
Base 25.0 HM 25.0
Lid 25.0 OBH 25.0
à
Set Point 37.0°C
Base 28.0 HM 29.3
Lid 31.2 OBH 30.2
Change the
Set Point
â
á
8
Confirm new
Set Point
Figure 20. Main Menu Display. Changing Set Points Description.
‘Main Page’ displays actual temperatures of Base, Lid, Objective Heater and Humidity Module
and the Set Point value.
Tip
►
The Objective Heater and humidity module temperatures are shown on display only if
they are active (See paragraphs 13.6 and 13.7)
Press ‘Enter’ to change the set point temperature.
‘Main Page’ can display temperature of a Free Sensor (Optional). Free Sensor can be used
to monitor the temperature inside the Petri dish, to calibrate the chamber and to check the
sample temperature during an experiment.
Figure 21 shows the ‘Main Page’ when working with Free Sensor connected to the Control Unit.
Set Point
37.0°C
Free Sensor
37.0°C
Base 37.0
HM 37.0
Lid 37.0
OBH 37.0
Figure 21. Main Page when working with Free Sensor connected to the Control Unit.
Selecting Offset Mode
13.5.1
Standard Offset Mode
When working in standard mode the temperatures of both the lid and the base of the
incubating chamber are strictly controlled. A careful calibration performed in our laboratories,
guarantees that sample temperature is maintained at the desired set point value.