32
SENSOR LID CODES
SENSOR LID-35
Temperature sensor lid. To be used in specimen feedback in a 35 mm Petri Dish
SENSOR LID-60
Temperature sensor lid. To be used in specimen feedback in a 60 mm Petri Dish
SENSOR LID-GS
Temperature sensor lid. To be used in specimen feedback in chamber slides and
chambered cover glass
Table 2. Sensor lid codes.
Figure 17. SENSOR LID-35. For 35 mm Petri-dish
It is important to monitor the room temperature during the calibration, as well as
reducing air drafts and forced convection. If possible divert any draft coming from
the air conditioning system in the room, from the incubation system.
See paragraph 13.5.2 for instructions on how to use the push button display to calibrate the
chamber.
From the main screen press the ‘Menu’ key on the instrument panel.
Press move the cursor using the up and down arrows to ‘Offset mode’ and
press ‘Enter’
Press ‘Enter’ with the cursor on ‘Mode’
Press ‘Up/down’ until you see ‘Custom’
Press ‘Enter’ to select ‘Custom’ as the configuration in use
Press the ‘Menu’ Key to exit from the Sub Menu.
Press ‘Menu’ key to go to the ‘Main Page’.
Tip
►
Reading Offsets can be varied manually to have the specimen at the set point
temperature.
Before calibrating the chamber manually, choose the Set Point Temperature
Press ‘Enter’ to change the Set point temperature to 37.0°C.
SENSOR LID
Fine Gauge
Thermocouple