38
Menu 1/3
à
Offset Mode
Offset
Objective Heater
8
Offset Mode
à
Mode
Standard
Ref @ 37°C: -1.3
Figure 22. Offset Mode Selecting.
Enter the Main Menu pressing the ‘Menu’ Key
Move the cursor to ‘Offset Mode’ and press ‘Enter’. This will show the current
base offset in Standard Working Mode.
Tip
►
The system is shipped from the factory already set according to the incubating chamber
model and sample adapter you have ordered and in standard working mode. You can read the
base offset in use on the display (See Figure 22). You can also see all the current offsets (lid
base and HM offsets) in ‘Offset’ submenu (See Figure 23).
Set Point 37.0°C
Base 37.0
HM 37.0
Lid 37.0
OBH 37.0
-
Menu 1/3
à
Offset Mode
Offset
Objective Heater
â
-
Go to main
Page
Menu 1/3
Offset Mode
à
Offset
Objective Heater
8
Current Offset
Base: -01.2
Lid: -03.2
HM: -03.0
Read Defaul Offset
Values
Figure 23. How to see the current offsets in standard mode