Equipment And Connections
Equipment Supplied
UNO includes:
1.
Control-Unit
2.
Humidifying Module Heater
3.
Glass Bottle - Humidifying Module (attention: FRAGILE) with TUBE E (140 mm
silicon tube of 6 mm OD with Filter).
4.
Pressure gauge with filter, regulator and assembly stirrup for CO
2
, degree of
filtration 20μm and condensate drain. Install pressure gauge between the pure CO
2
tank
pressure regulator and the Control Unit input port.
5.
Power cord cable (x1). Use power cord with Control Unit. The UNO-COMBINED-
CONTROLLER can be powered at a voltage within the range 110-230VAC
6.
TUBE E. 3 m long tube assembled as follows: 2 m long rigid tube of 6mm OD,
connected to 0.5m long 4mm ID silicon transparent tube. This assembly is then
connected to an hydrophobic filter 0.2 micron porous and to a 0.5 m long 4mm ID
silicon transparent tube. Connect one end of this tubing assembly to the output port
labeled “Air+CO
2
out
”
and the other end to the Humidity Module.
7.
TUBE H. 1.5 m long tube composed of 4mm ID silicon transparent tube (0.5 m long)
connected to 2mm ID silicon transparent tube (1 m long) (both wrapped in an insulating
material). Use insulated tubing in point 8 to connect the Humidity module to the
Incubating chamber.
The following items are required but sold separately.
‘UNO’ must be equipped with:
8.
Incubating chamber (model depending on the XY stage with or without z piezo) with
heated lid, chamber riser and one or more plate adapters.
9.
Flat screwdriver or Allen wrench. To screw the grub screws or the screws on the
sides of the incubating chamber plugging the built in perfusion holes.
Optional accessories:
OKO-AP + Power cord (x1). The OKO-AP can be powered at a voltage within the
range 110-230VAC
PRESSURE GAUGE WITH FILTER: When a compressed air line is available in the lab
it is possible to avoid using an air pump such as OKO-AP. In this case a Pressure gauge
with filter, regulator and assembly stirrup, degree of filtration 20μm and condensate
drain, for Air is available upon request.
Objective heater. Required ONLY when using oil immersion objectives.