3/Checkout
3-1 3-1
3/Checkout
In this section
Mechanical checks .............................................................................................................................................3-1
Controller checks .................................................................................................................................................3-1
Operational checks .............................................................................................................................................3-2
WARNINGS
Do not perform the preoperative checkout procedure while a patient occupies
the incubator.
Complete the “Checkout ” section of this manual before putting the unit into
operation. If the servo-controlled humidifier fails any portion of the checkout
procedure it must be removed from use and repaired.
Mechanical checks
WARNING
Disconnect the power cord for the mechanical portion of the preoperative
checkout procedure.
1. Disconnect the servo-controlled humidifier power cord.
2. Examine the power cord for damage. If the cord is damaged, replace it.
3. Examine the servo-controlled humidifier for obvious signs of damage.
4. Check that the humidifier controller is securely fastened to the incubator base.
5. Locate the humidifier fill tray on the underside of the base. Slide the tray out. Verify
that the seals are in good condition and that there is no obvious damage to any
of the humidifier components. Inspect the fill tray electrical connector for damage.
Inspect the tray for cracks or crazing. Replace any damaged parts. Slide the
humidifier tray back in. Push it back until you feel slight resistance, then push until
it snaps into position.
Controller checks
WARNING
Do not use the Care Plus Servo-controlled Humidifier in the presence of
flammable anesthetics; an explosion hazard exists under these conditions.
1. Make sure the power cord is connected to the inlet socket on the rear of the
humidifier controller.
Содержание Ohio Care Plus
Страница 20: ...2 General Information 2 4 2 4 Notes...
Страница 24: ...4 Using the Servo controlled Humidifier 4 2 4 1 Notes...
Страница 46: ...6 Service 6 18 6 18...
Страница 53: ...8 Schematics 8 1 Figure 8 1 Wiring Diagram Servo controlled Humidifier 8 Schematics CI 18 059...
Страница 54: ...8 Schematics 8 2 Figure 8 2 Power Supply section of the Control Display Board Assembly CI 18 060...
Страница 55: ...8 Schematics 8 3 Figure 8 3 Digital section of the Control Display Board Assembly CI 18 061...
Страница 56: ...8 Schematics 8 4 Figure 8 4 Analog section of the Control Display Board Assembly CI 18 062...
Страница 57: ...8 Schematics 8 5 Figure 8 5 Heater Control section of the Control Display Board Assembly CI 18 063...
Страница 58: ...8 Schematics 8 6 Figure 8 6 Display section of the Control Display Board Assembly CI 18 064...
Страница 67: ......