B-2
APPENDIX B SERVICE CALIBRATION
10. When calibration is complete, the weight is indicated in the first window, center and right windows
indicate 0. See Figure B-5.
Figure B-5. Final Calibration Weight Display.
11. Remove all calibration weights from the scale
B.1 CALIBRATION (Cont.)
9. Add the third weight to the total indicated on the scale (3000g shown), then wait. See Figure
B-4.
Figure B-4. Third Calibration Weight Display.
Содержание EC Series
Страница 2: ......
Страница 4: ......
Страница 14: ...2 6 CHAPTER 2 DIAGNOSIS...
Страница 30: ...3 16 CHAPTER 3 SCALE TESTING AND CALIBRATION...
Страница 52: ...A 4 APPENDIX A SERVICE MODES...
Страница 55: ......
Страница 56: ...PN 80010584 B SERVICE MANUAL EC SERIES SCALES 80010584...