3-11
CHAPTER 3 SCALE TESTING AND CALIBRATION
3.6.4 Linearity Test
The linearity test is used to determine the linearity of the scale throughout its operating range. The results
should be in accordance with the specifications shown above. The reference mass should be approxi-
mately 1/8 of the full-capacity of the scale.
NOTE
:
: :
:
: The scale must pass the Off-Center Load test and Repeatability Test
before
the Linearity Test is
performed.
BE CERTAIN TO USE THE SAME REFERENCE MASS THROUGHOUT THE LINEARITY TEST.
1. Press
ZERO
to zero the display.
2. Place the reference mass on the center of the platform and record the exact value (all decimal
places) displayed.
3. Remove the reference mass and verify that the display returns to zero. If it does not, disregard the
reading, press
ZERO
to rezero the display, and take the reading again.
4. Place a mass equal to 1/4 of the full scale capacity on the platform and press
ZERO
.
5. Add the reference mass to the center of the platform and record the exact value displayed.
6. Remove the reference mass and verify that the display returns to zero. If it does not, disregard the
reading, press
ZERO
to rezero the display, and take the reading again.
7. Place a mass equal to 1/2 of the full-scale capacity on the platform and press
ZERO
.
8. Add the reference mass to the center of the platform and record the exact value displayed.
9. Remove the reference mass and verify that the display returns to zero. If it does not, disregard the
reading, press
ZERO
to rezero the display, and take the reading again.
10. Place a mass equal to 3/4 of the full scale capacity on the platform and press
ZERO
.
11. Add the reference mass to the center of the platform and record the exact value displayed.
12 Remove the reference mass and verify that the display returns to zero. If it does not, disregard the
reading, press
ZERO
to rezero the display, and take the reading again.
13. Calculate the weight differences between the reading as shown in the recorded readings.
14. The Maximum difference should be less than or equal to the Linearity specification for the
scale.