3-12
CHAPTER 3 SCALE TESTING AND CALIBRATION
3.6.4 Linearity Test (cont.)
Record Readings
Reading 1 reference weight added to zero
________________
Reading 2 reference weight added at 1/4 full-scale capacity ________________
Reading 3 reference weight added at 1/2 full-scale capacity ________________
Reading 4 reference weight added at 3/4 full-scale capacity ________________
Calculate Differences
Reading 1 – [(Reading 1 + Reading 2) / 2] =
________________
Reading 2 – [(Reading 2 + Reading 3) / 2] =
________________
Reading 3 – [(Reading 3 + Reading 4) / 2] =
________________
Maximum Difference
_____________
Содержание EC Series
Страница 2: ......
Страница 4: ......
Страница 14: ...2 6 CHAPTER 2 DIAGNOSIS...
Страница 30: ...3 16 CHAPTER 3 SCALE TESTING AND CALIBRATION...
Страница 52: ...A 4 APPENDIX A SERVICE MODES...
Страница 55: ......
Страница 56: ...PN 80010584 B SERVICE MANUAL EC SERIES SCALES 80010584...