EN-18
Adventurer SL
CAL is displayed. Press
Yes
to enter the
Calibration Sub-menu.
When SPAN displays, press
No
to display
LINEAR, the next Menu Item.
Press
Yes
to initiate Linearity Calibration.
The display flashes while the zero reading
is taken.
The display flashes the midpoint weight
value for Linearity Calibration, according
to the specific model.
Put the specified weight on the pan and
press
Yes
to calibrate.
The display flashes BUSY.
The display flashes the capacity weight
value, according to the specific model.
Put the specified weight on the pan and
press
Yes
to calibrate.
The display flashes BUSY.
When calibration is complete, DONE is
displayed. Remove the weight from the
pan. The balance returns to the currently-
selected application mode.
3.5.1.2 Linearity Calibration (Cont.)