Adventurer SL
EN-17
When SPAN displays, press
Yes
to initiate
Span Calibration, or
No
to go to Linearity
Calibration.
The display flashes while the zero reading
is taken.
The display flashes the default maximum
weight for the specific model. Press
No
to
select an alternate calibration weight. Span
calibration weight values are shown in
Section 5.4.
Put the indicated calibration mass on the
pan and press
Yes
to calibrate.
The display flashes BUSY.
When calibration is complete, DONE is
displayed. Remove the weight from the pan.
The balance returns to the currently selected
mode.
3.5.1.2 Linearity Calibration
Linearity Calibration is an advanced method used to reset the balance's weighing
range using three weight values:
• zero
• a weight value at or near the midpoint of the balance's weighing range
• a weight at or near the specified capacity of the balance.
Linearity calibration is required if testing shows that the linearity error exceeds the
Linearity tolerance in the Specification table.
3.5.1.1 Span Calibration (Cont.)
To perform Linearity Calibration:
With the balance turned off, press and
hold
On
until MENU is displayed, then
release the button.