Clamp pincer CP 10 / CP 20 / CC 20
5
Use of clamp pincer
2. In the pop-up menu, click on OK to confirm.
Initial calibration
The “Calibration/Initial-calibration” window now
opens.
The initial closing force has already been input
and doesn’t need to be re-entered.
3. Click on the “Start initial calibration” button.
The “Initial calibration” information window
opens. This describes the next steps and
contains a counter showing the number of
measurement closures.
Once the measurement closure has been
performed the “Initial calibration” window
closes automatically.
4. Input the measured value into the
“Measuring value [N]” field.
The “Send measuring value to device” button
lights up green.
5. Click on the “Send measuring value to
device” button.
The measuring value is acknowledged and
sent to the device.