18 Cleaning, Di
s
infection and Maintenance
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
71 / 92
Caution
Risk of personal injury or material damage due to invisible laser radiation
The Myopia Master® contains a Class 1 laser according to DIN EN 60825-
1:2015 and DIN EN 60825-1: 2001. It is an encapsulated laser system.
When the Myopia Master® cover is opened, you may be exposed to
invisible, Class 3R (5 mW) laser radiation.
Never open the unit.
For authorized service personnel only: When doing maintenance
jobs, avoid looking directly into the laser beam.
18.4 Attaching Paper to the Chin Re
s
t
If you want to attach new chin rest paper, follow these instructions:
Pull the two pins (1) out of the chin rest.
Put the chin rest paper (2) in such a way that the holes of the paper
and the chin rest (3) are aligned.
Insert the two pins (1) in the chin rest.
18.5 In
s
erting a New Roll of Printer Paper
Caution
Risk of electric shock if the Myopia Master® is not completely
disconnected from the mains
Turn the Myopia Master® off,
Pull the power plug before inserting a printer paper. When
disconnecting electrical connections, pull on the respective plug and
not on the cable itself.
1
Pins
3
Chin rest
2
Chin rest paper
Fig. 18-2: Attaching chin rest paper
1
3
2