Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
Do not use excessive force when connecting the electrical plug.
If a connection is not possible, check whether the plug fits the socket.
If you find damage to the plug connector, have the damage
corrected by our service department.
Establish an USB connection only with the OCULUS USB FS MED-
Isolator (Nr. 01 56920 00 010).
Note that an output voltage of maximum 5.5 V DC is supplied by a
device connected via USB.
Do not use the Myopia Master® with wireless technology, for
example with wireless USB.
Data responsibility:
The device itself is not designed to connect with
the internet, but only to a PC. It does not require the internet to
function.
Do not connect with the internet while using the device. It is consid-
ered misuse.
If you elect to connect the PC to the internet for other purposes you
are responsible for ensuring data security.
Only use a Myopia Master® which is mounted properly.
Patient environment information
Patient environment is the area where patients can come into contact
with any part of a medical electrical equipment (ME equipment) or with
another person being in contact with the ME equipment.