14 Reference Mea
s
urement
50 / 92
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
14 Reference Mea
s
urement
To achieve a high measuring accuracy, the Myopia Master® must be set up
before conducting the first examination on a patient
after changing the position of the Myopia Master®
The first reference measurement is performed during setup by OCULUS or
an authorized dealer. OCULUS recommends performing a reference
measurement once each month.
The reference measurement can be performed easily and quickly using the
reference tool.
Required materials
reference tool, provided
cleaning agent, see
Measuring with the reference tool
Prerequisite: the Myopia Master® must be turned on for at least 15
minutes. For the reference measurement, proceed as follows:
Remove the cover cap.
Thoroughly clean the reference tool before saving reference values
with the cleaning agent.
Place the reference tool on the chin rest.
Add a new patient, named "reference test" and select "Myopia" or
"ARK + AXL".
Perform a measuring operation with the reference tool
.
Compare the results with the results on the reference tool.
Fig. 14-1: Installation of the reference tool