Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
9 / 92
Ensure that connections with non-medical devices are made
correctly.
Only use the power adapter listed in the packing list.
Establish an USB connection only with the OCULUS USB FS MED-
Isolator (Nr. 01 56920 00 010).
Use only a computer that meets the specifications given in this
instruction manual,
Note that an output voltage of maximum 5.5 V DC is supplied by a
device connected via USB.
Caution
Use of a multiple socket outlet
Risk of personal injury or material damage caused by unsafe multiple
socket outlet
If you use a multiple socket outlet to connect the
Myopia Master® to the power supply, you must heed the following
information:
Use an extension cord that complies with the requirements of
IEC 60601-1: 2005, section 16.
Do not place the multiple socket outlet on the floor.
Do not use more than one multiple socket outlet.
Plug only the Myopia Master® and the computer that is being used
with the unit (if applicable) into the multiple socket outlet.
If you are using a multiple socket outlet it has to be supplied with a
isolation transformer.
If you are using a new computer for the Myopia Master®, you must have
the electrical safety checked. Call OCULUS Service for this purpose.
Electromagnetic Compatibility (EMC) / Cables
Risk of personal injury or damage to property due to electromagnetic
interference
Portable and mobile RF communications equipment can affect medical
electrical equipment
.
Make sure that portable and mobile RF communications equipment
do not cause interference.
Recommendation: Maintain a minimum distance of 4 m. If the
distance is shorter, you must ensure that the Myopia Master®
functions correctly.