17 Calculated correction of tonometrically mea
s
ured IOP
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
65 / 92
17.4 Performing IOP correction with the Myopia Ma
s
ter®
Select an IOP-Table in the menu "Settings 1"
.
Perform a P+AR+K or a PARK + AXL measurement
.
The results screen appears for the examined eye:
The IOP change value is displayed in the IOP change field (1). Depending
on the formula which is used, the IOP change value is either a summand
or a factor.
Turn the control wheel until the "IOP Chg." field is activated.
Press the control wheel.
The field for entering the tonometrically measured IOP value appears:
1
IOP
change field
Fig. 17-3: IOP change
Fig. 17-4: Entering tonometrically measured IOP manually
1