7.10.4 Cabinet Reset
The main control board has two software operating resets available for qualified service personnel. The two types
are the following:
Factory Reset: Resets setpoints and selected option settings. Factory reset should be used in the event the
system memory develops an error in operation. Cabinet type, motor type, and calibration data
will not be affected with this reset.
Master Reset: Resets all calibration, cabinet type, motor type sensor data, and options settings back to default
settings. Master reset should only need to be used for a main control board replacement.
After pressing the Calibration/Service menu item, a password screen will appear. The default password is “9876”.
Once the service password is entered, the Calibration/Service menu will appear.
•
Select CABINET RESET from the menu.
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...