11.0 Electrical/Environmental Requirements
11.1 Electrical
(Supply Voltage Fluctuations Not to /- 10%)
*NU-581-400
115 VAC,
60 Hz,
1 Phase,
14 Amps
*NU-581-500
115 VAC,
60 Hz,
1 Phase,
16 Amps
*NU-581-600
115 VAC,
60 Hz,
1 Phase,
16 Amps
*UL/UL-C Listed
11.2 Operational Performance
(for indoor use only)
Environment Temperature Range:
60
°
F-85
°
F (15
°
C - 30
°
C)
Environment Humidity:
Maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 50% relative humidity at 40°C
Environment Altitude:
6562 Feet (2000 meters) maximum
11.3 Light Exposure
Standard LED Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
11.4 Installation Category:
2.0
Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to
withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For example,
in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public mains
such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is 2500 V
for a 230 V supply and 1500 V for a 120 V supply.
11.5 Pollution Degree:
2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2
assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity
caused by condensation.
11.6 Chemical Exposure
Chemical exposure should be limited to antibacterial surface disinfectants used for cleaning and disinfecting.
USE OF
CHLORINATED OR HALOGEN MATERIALS IN THE CABINET MAY DAMAGE STAINLESS STEEL.
Equipment decontamination can be accomplished by non-condensing gas or vapor Paraformaldehyde, Hydrogen
Peroxide or Chlorine Oxide following NSF/ANSI 49, Annex I-2 (formerly Annex G).
11.7 EMC Performance (classified for light industrial)
Emissions:
EN61326
Immunity:
EN61326
Class A equipment is intended for use in an industrial environment. In the
documentation for the user, a statement shall be included drawing attention to
the fact that there may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated
disturbances.
WARNING
!
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...