OM0302
Rev 2/December 2020
Page 86 of 107
o
Blower Window Interlock
This parameter allows for the selection of the window closed switch to be interlocked to the blower.
When the blower interlock is on, the blower will turn off when the window is closed. When the blower
is off, the blower will continue to run when the window is closed.
o
Fan Relay Interlock
This parameter allows for the selection of the fan relay interlock operation. When the fans relay
interlock is on, and the blower switch is pending or blue, the fan relay will be off or not energized. If the
fan relay interlock is off and the blower switch is pending or blue, the fan relay will be on or energized.
In either case the fan relay will be on when the blower switch is on or green and off when the blower
switch is off or not colored.
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...