OM0302
Rev 2/December 2020
Page 85 of 107
All interlocks are enable/ disabled by a user through the interlock Features screen
o
UV window interlock:
This parameter allows for the selection of the window closed switch to be interlocked with the UV light
option. When the window interlock is on, the window must be closed for the UV light to operate.
When the window interlock is off the UV light can be turned on regardless of the window position.
NOTE
: In addition to the
TOUCHLINK
system UV window interlock there is a double redundant UV
window
interlock relay. To override the UV window interlock for service purposes only, both interlocks must be
changed through the
TOUCHLINK
system and shorting the relay connection. (See electrical schematic
for reference)
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...