5.2.5 Plumbing Services
Service ball valves with the type of service specified by the removable button on the handle are located in the
work zone. The service ball valves are not recommended for pressure over 75 p.s.i. (5.2 BAR). Reducing valves
should be installed external to the cabinet if necessary. Service ball valves should never be used for flammable
gasses or oxygen service. A special needle valve for oxygen service or certified valve is required and available upon
request.
External connection is to 3/8 inch NPT coupling in the inner sidewalls. Connection to plant utilities should be made
with proper materials for the individual service and according to national and/or local codes. Observe all labels
pertaining to the type of service and operating pressure.
5.2.6 Electrical Services
The NU-581 series Biosafety Cabinets may be "hardwired" (optional) or plugged into an outlet with
protective earthing connection with the standard power cord. The cabinet requires 115 VAC, 60 Hz, single phase
(current rating varies per cabinet size, reference Electrical/Environmental Requirements). It is recommended that
power to the cabinet, whether hardwired or plug connected, be on its own branch circuit, protected with a circuit
breaker at the distribution panel near the cabinet.
A surge protector is strongly recommended
if you are
experiencing power related faults.
NOTE: THIS CABINET CONTAINS ELECTRONIC BALLASTS FOR THE
LED
LIGHTING. ELECTRONIC BALLASTS
OPERATE WITH HIGH INRUSH CURRENT. IT IS NOT RECOMMENDED TO USE THIS PRODUCT WITH
GROUND FAULT CIRCUIT INTERRUPTERS (GFCI’S) BECAUSE THE BALLASTS MAY CAUSE THE GFCI TO
TRIP.
5.2.7 Final Assembly
NOTE: Remove the protective cardboard cover over the exhaust HEPA filter
.
Remove the protective cardboard cover over the exhaust HEPA filter located under the protective screen if in
place. The exterior surface and viewing glass are easily cleaned with any mild household detergent cleaner using a
soft cloth. Harsh chemicals, solvent-type cleaners and abrasive cleaners should not be used.
Do not attempt to clean the HEPA filter media. Cabinet interior walls or work surface are easily cleaned with any
mild household detergent cleaner using a soft cloth. Turn the cabinet on and let it operate for 60 minutes before
using it as a LFBSC.
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...