OM0302
Rev 2/December 2020
Page 50 of 107
8. Remove bagged prefilter from cabinet and place in appropriate transport container.
9.
Before installing new HEPA prefilter, clean surface of filter gasket with 70% IPA.
It is also recommended to apply a very thin layer of silicone grease on the gasket.
If gasket is damaged (i.e. torn), replace it (Nuaire part number A-3071-25).
Proceed to install HEPA prefilter making sure filter is seated level onto the gasket.
NOTE:
Only use same type of HEPA prefilter to assure proper function of the cabinet.
10. Turn to engage stainless hold-down clamps (2) and replace protective screen.
11. Repeat the process for each HEPA prefilter as needed until complete. (See section 7.6.1 for HEPA prefilter testing).
Hazardous Material!
Follow all Local, State and Federal Guidelines for disposal.
WARNING
!
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...