OM0302
Rev 2/December 2020
Page 43 of 107
Determine decontamination system cycle parameters.
Note that system parameters will change for process type, as well as cabinet model and size.
For validation purposes, a LabGard® ES Model NU-543-600E was set using the following parameters on the
Bioquell Clarus II:
Pressure Setpoint:
-10 Pa
Airflow Setpoint:
22 m₃/hr
Delivery Temperature:
65⁰C
Conditioning Time:
10 min
Gassing Time:
40 min
Gassing Injection Rate: 6.0 g/min
Dwell Time:
90 min
Injection in Dwell:
ON
Dwell Injection Rate:
3.5 g/min
Aeration Delay:
60 min
Aeration Time:
600 min
The following process was performed by the Steris X10 using the LabGard ES Model NU-543-400
1.
Blower is turned off.
2.
Biological Indicators were placed at the exhaust filter, corner of working area and under the work tray of
BSC or anywhere inside the cabinet.
3.
Exhaust opening sealed with a metal plate and a tubing connection for the Hydrogen Peroxide (H
2
O
2
)
return.
4.
Front opening sealed with plastic film and a tubing connector mounting plate. Reference BCD-16509.
5.
Edges of the front window sealed with a low residue tape.
6.
Tubing connected to the inlet and outlet of X10.
7.
Insertion of Hydrogen Peroxide (H
2
O
2
) sterilant cup and desiccant cartridge to X10.
8.
Current BSC process selected.
Decontamination Process:
9.
The X10 will perform one (1) Preheat, Dehumidification, Decontamination and Aeration after start of the
program.
10.
Preheat –
The cabinet is heated to 31⁰ C.
11.
Dehumidification – Cabinet is dehumidified to 15% RH.
12.
Condition – Sterilant vapor is added to sealed cabinet to target concentration.
13.
Checked for any Hydrogen Peroxide (H
2
O
2
) leakage from the cabinet. Applied extra tape as needed.
14.
Decontamination – The cabinet is filled with sterilant vapor for 45 minutes.
15.
During the decontamination, the blower was activated for 1 minute every 15 minutes.
Aeration:
16.
Aeration – The sterilant vapor is converted to water and oxygen by 3 options
a.
Connect the exhaust to Fumehood or outdoors
b.
Connect the AR60 Aerator which catalyzes the Hydrogen Peroxide (H
2
O
2
) vapor about 3 hours.
c.
Run Aeration cycle of the X10 for about 10 hours.
17.
Check the Hydrogen Peroxide (H
2
O
2
) level during the Aeration cycle and open the seal when level is less
than 1ppm.
Biological indicators were removed and incubated for 7 days.
Содержание LabGard NU-581-400
Страница 7: ...OM0302 Rev 2 December 2020 Page 7 of 107...
Страница 9: ...OM0302 Rev 2 December 2020 Page 9 of 107...
Страница 10: ...OM0302 Rev 2 December 2020 Page 10 of 107...
Страница 11: ...OM0302 Rev 2 December 2020 Page 11 of 107...
Страница 29: ...OM0302 Rev 2 December 2020 Page 29 of 107 Set New Password Retype new password...
Страница 42: ...OM0302 Rev 2 December 2020 Page 42 of 107...
Страница 100: ...OM0302 Rev 2 December 2020 Page 100 of 107...
Страница 101: ...OM0302 Rev 2 December 2020 Page 101 of 107...
Страница 102: ...OM0302 Rev 2 December 2020 Page 102 of 107...
Страница 103: ...OM0302 Rev 2 December 2020 Page 103 of 107...
Страница 104: ...OM0302 Rev 2 December 2020 Page 104 of 107...
Страница 105: ...OM0302 Rev 2 December 2020 Page 105 of 107...
Страница 106: ...OM0302 Rev 2 December 2020 Page 106 of 107...
Страница 107: ...OM0302 Rev 2 December 2020 Page 107 of 107...