Rev. 01
Model 2001 Service Manual
35
Section 8
Maintenance
General
This section presents recommended maintenance schedules for the Model 2001 and
information on general maintenance, such as battery and fuse replacement, disassembly and
assembly instructions, and system software updates.
Maintenance Schedules
The electronic circuits within the Model 2001 Pulse Oximeter monitor do not require scheduled
calibration or service. However, in order to maximize battery life, the monitor’s internal battery
should be exercised monthly. Novametrix recommends the following maintenance schedules.
*
•
Cleaning and Sterilization:
Perform as required. See "Cleaning and Sterilization" on page 35
•
Battery Life and Maintenance:
See "Battery Life and Maintenance" on page 37
•
Functional Test:
This test verifies overall functional integrity of the monitor and sensors. See "Functional
•
Accuracy Test:
This test verifies the calibration accuracy of the monitor using specified test apparatus.
See "Accuracy Test" on page 27.
•
Electronic Tests:
These tests contain information on testing the electronic circuits within the Model 2001
and should only be performed if the monitor fails to pass the Functional Tests. Only
qualified service personnel should attempt to perform the Electronic Tests. See
"Electronic Tests" on page 29.
Cleaning and Sterilization
Follow the cleaning and sterilization instructions listed below to clean and/or sterilize the
monitor and its accessories.
*At the customer’s request, Novametrix will provide repair and calibration services under the terms of a Service
Contract. Contact the Novametrix Service Department for contract details.
Содержание MARS PO2 TECH 2001
Страница 2: ......
Страница 4: ...Manufacturing Quality and Safety iv Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 10: ...Section 1 Patient Safety Cautions 4 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 40: ...Section 7 Electronic Tests Safety Testing 34 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......