Section 2
Introduction
Indications and Usage
6
Model 2001 Service Manual
Rev. 01
Model 2001 is calibrated to display “functional” saturation. This differs from the “fractional”
saturation value displayed by most co-oximeters.
Functional saturation represents the amount of oxyhemoglobin as a percentage of the
hemoglobin that can be oxygenated. Dysfunctional hemoglobins (COHb and METHb) are not
included in the measurement of functional saturation.
Pulse Rate is calculated by measuring the time interval between the peaks of the infrared light
waveform. The inverse of this measurement is displayed as pulse rate.
Model 2001 must be used in conjunction with SuperBright™ saturation sensors. (An CHECK
SENSOR SITE display message indicates a non-SuperBright™ Sensor may be in use.)
Indications and Usage
The Model 2001 Pulse Oximeter Monitor is intended to be used for monitoring oxygen
saturation and pulse rate in all critical monitoring environments including ventilatory support
and anesthesia. Model 2001 is designed to monitor all patient areas including adult, pediatric
and neonatal.
Symbols
Equation 1. Functional
Saturation Calculation
Functional Saturation =
HbO
2
100 - (COHb + METHb)
HbO
2
= Fractional Hemoglobin
COHb = Carboxyhemoglobin
METHb = Methemoglobin
Symbol
Description
Patient Isolation
Identifies patient isolation connection as type BF.
Attention
Consult manual for detailed information.
Separate collection
Ensure that spent batteries are collected separately when
disposed of. Found on the internal battery. Refer to quali-
fied service personnel when battery replacement is
required.
Recyclable item
Found on the internal battery. Refer to qualified service
personnel when battery replacement is required.
Indicates heavy metal content, specifically lead. Found
on the internal battery and monitor enclosure. Refer to
qualified service personnel when battery replacement is
required.
!
Pb
Содержание MARS PO2 TECH 2001
Страница 2: ......
Страница 4: ...Manufacturing Quality and Safety iv Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 10: ...Section 1 Patient Safety Cautions 4 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 40: ...Section 7 Electronic Tests Safety Testing 34 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......