Rev. 01
Model 2001 Service Manual
45
Section 10
Specifications
General
Specifications for the Novametrix
Model 2001
Pulse Oximeter, are listed for informational
purposes only, and are subject to change without notice.
Oxygen Saturation (SpO
2
) Section
•
Range: 0-100%
•
Accuracy: 70-100%
±
2% (1 standard deviation), 0-69% unspecified
(Approximately 68% of the observations are within the accuracy claim.)
•
Display Resolution: 1%
•
Averaging Time: fixed at 8 seconds
•
Audible SpO
2
Trend Feature:
Pitch of (user selectable) Pulse Rate “beep” tracks the SpO
2
value (i.e., decreasing SpO
2
values are signalled by lower pitched “beeps”).
•
Settling Time:
Display settles to within 1% of the final reading less than 15 seconds after the sensor is
properly applied.
•
Alerts:
Continuously displayed. Menu selectable high and low limits (100-50). Visible alarm is
immediate. Audible alarm occurs after 10 seconds of continuous violation of the set limit,
or immediately. Limit values are retained in memory when monitor is turned off, or the
monitor can be set to use its default settings each time it is turned on.
Pulse Rate Section
•
Range: 30-250 beats per minute (bpm)
•
Accuracy
: ±
1% of full scale (1 standard deviation)
(Approximately 68% of the observations are within the accuracy claim.)
•
Display Resolution: 1 bpm
•
Averaging Time: fixed at 8 seconds
•
Settling Time:
Display settles to within 1% of the final reading less than 15 seconds after the sensor is
properly applied.
•
Alerts:
Continuously displayed. Menu selectable high and low limits (249-30 or Off). Visible alarm
is immediate. Audible alarm occurs after 10 seconds of continuous violation of the set limit
or immediately. Limit values are retained in memory when monitor is turned off, or the
monitor can be set to use its default settings each time it is turned on.
Содержание MARS PO2 TECH 2001
Страница 2: ......
Страница 4: ...Manufacturing Quality and Safety iv Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 10: ...Section 1 Patient Safety Cautions 4 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 40: ...Section 7 Electronic Tests Safety Testing 34 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......