Section 9
Troubleshooting
Fault, alert and error messages that can appear on the Model 2001 Message Center display
are described below.
Fault and Error Condition Messages
SENSOR OFF PATIENT
Sensor disconnected from patient, improperly applied, or placed on
an area too translucent for proper sensor operation. Reposition
sensor.
BAD SIGNAL TIMEOUT
Monitor not receiving valid signals from sensor. May be caused by
excessive motion, cardiac arrhythmia or other situations leading to
poor signal. Check patient status, reposition sensor. Changes to
PULSE SEARCH after 30 seconds.
CONNECT SpO2 SENSOR 1. Sensor is disconnected from the monitor.
2. Sensor is faulty. Remove sensor from use and contact qualified
service personnel.
3. Sensor is placed on a site too thick. Reposition the sensor on a
thinner (less opaque) section of tissue.
ERROR - FAULTY SEN-
SOR
Sensor faulty. Remove sensor from use and contact qualified service
personnel.
CHECK SENSOR SITE
Sensor placed on a site too thick (or opaque) for adequate light
transmission.
1. A non-SuperBright™ sensor is connected, use only 87xx series
sensors.
2. Sensor is faulty. Remove sensor from use and contact qualified
service personnel.
SHIELD SENSOR
HIGH AMBIENT LIGHT
Ambient light sources (sunlight, warming lights, etc.) are interfering
with sensor light sources. Shield the sensor from ambient light
sources. Changes to REPOSITION SENSOR after 30 seconds.
LOW SIGNAL STRENGTH
Pulse strength as detected by sensor is too weak for proper monitor
operation. Reposition sensor. Changes to REPOSITION SENSOR
after 30 seconds.
MONITOR ERROR *
Monitor faulty, where * is a message or error code. Record error
message (appearing on bottom line of display) and contact qualified
service personnel.
Содержание MARS PO2 TECH 2001
Страница 2: ......
Страница 4: ...Manufacturing Quality and Safety iv Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 10: ...Section 1 Patient Safety Cautions 4 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 40: ...Section 7 Electronic Tests Safety Testing 34 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......