Battery Life and Maintenance
Section 8
Maintenance
Rev. 01
Model 2001 Service Manual
37
Battery Life and Maintenance
The monitor was shipped from the factory with a fully charged battery. Since the monitor draws
some battery power even while turned off, it is recommended that the monitor be operated on
line power for the first 24 hours to allow ample time for the battery to fully recharge.
The monitor may be operated on line power while the battery is recharging. Approximately 16
hours are required to fully recharge a discharged battery.
During prolonged periods of storage or shipment, the battery may discharge enough to prevent
the unit from turning on while operated from battery power. If this occurs, plug in the line cord,
set the switch on the rear panel to the “|” On position, ensure that the green
indicator on the
front panel is illuminated, and allow the monitor to charge for 24 hours before switching it on.
Mains Voltage Configuration
The rear panel power entry module indicates the mains voltage setting for the monitor. Check
that the voltage is correct before attaching the AC line cord and powering the monitor. The
Model 2001 can be set to operate from 100-120 VAC 50/60Hz or 200-240VAC 50/60Hz.
Instructions for fuse replacement and changing the mains voltage setting follow.
Fuse Replacement
1.
Check that the monitor is OFF.
2.
Set the rear panel power entry module switch to OFF (“O”). Remove the AC line cord from the
power entry module.
3.
Using a flat blade screwdriver, pry the fuse access door open to expose the fuse housing. Note
the orientation of the fuse housing (this determines the mains operating voltage).
CAUTION:
Replace fuses with same type and rating. Verify proper fuse value for mains voltage setting
(see table below)
.
Mains Voltage
Fuses (Slo Blo)
100-120 VAC
0.5 A 250V
200-240 VAC
250mA 250V
Fuse housing
Fuse access door
Power entry module
Содержание MARS PO2 TECH 2001
Страница 2: ......
Страница 4: ...Manufacturing Quality and Safety iv Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 10: ...Section 1 Patient Safety Cautions 4 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 40: ...Section 7 Electronic Tests Safety Testing 34 Model 2001 Service Manual Rev 01 This page intentionally blank...
Страница 66: ...Section 12 Parts Lists 60 Model 2001 Service Manual Rev 01...
Страница 68: ......
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 74: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......
Страница 81: ......
Страница 82: ......
Страница 83: ......
Страница 84: ......
Страница 85: ......