6.5.2 Adjusting the Cooling/Coaxial, Valve, and Fluid Air Pressure
1.
Locate the Cooling/Coaxial, Valve, and Fluid Pressure Regulator adjustment knobs in the
front cabinet (Figure 6-4).
TIP
For accurate pressure adjustment, lower the pressure below the desired level and
then increase to the desired pressure.
2.
Rotate the adjustment knob counterclockwise until the gauge registers 0 psi.
3.
Rotate the adjustment knob clockwise until the associated gauge registers the air pressure
value (psi or kPa) required by your dispensing application.
"
NOTE
Pressure settings depend on the fluid being dispensed and dispensing valve being
used. Refer to the applicable dispensing valve manual for additional information.
4.
Monitor the pressure gauge on the front panel to make sure that pressure builds at a steady
rate. A drop in air pressure can indicate an air leak.
!
If there is an air leak, identify the source, shut off the facility air, and fix the leak before
proceeding.
Item Description Item Description
1
Vacuum Control Regulator
3
Valve Air Regulator
2
Cooling/Coaxial Air Regulator
4
Fluid Air Regulator
Figure 6-4 Adjusting the Valve and Fluid Pressure Regulators
1
2
3
4
Содержание Asymtek Spectrum S-920
Страница 1: ...Spectrum Series S 920 Dispensing System Installation Operations Maintenance Manual P N 7212417 Rev A...
Страница 2: ......
Страница 4: ......
Страница 30: ...1 14 Introduction 1 10 5 Rear View Figure 1 7A Rear View 2 3 4 1 5...
Страница 124: ......
Страница 172: ...7 12 Maintenance Figure 7 7 Lubricating the X Axis Linear Guides Figure 7 8 Lubricating Y Axis Linear Guides...
Страница 185: ...Troubleshooting 8 11 Figure 8 3 FmXP Scale Setup Menu Figure 8 4 Scale Error Message...
Страница 190: ......
Страница 204: ......
Страница 208: ......
Страница 210: ......
Страница 220: ......
Страница 222: ......
Страница 223: ......
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 234: ......