63
Calibrate (via RS232)
This option is for generation of certificates via the RS232 interface.
Ensure the Calibration Options are set up correctly.
Calibrate (via RS232) is available for all tool types.
1. Select the tool to be calibrated, then select
Calibrate (via RS232)
from the right mouse button menu
2. Select or Enter an
Inspector
(this must be entered)
.
3. Enter
Certificate Number
(if required).
4. Select
Direction
required
5. Enter the
Temperature
(in degrees C)
that the calibration was performed
(this must be entered)
.
6. Enter
Certificate
Text
Choose either
As
Found
or
As
Left
from the drop down
or enter your own text.
7. If the same tool has other calibration data to put on the same certificate, press
Select Results
This feature allows clockwise & anti-clockwise as well as ‘as found’ & ‘as left’ results to be included on
the same certificate.
To add another calibration to the same certificate double click on the required results.
8. Select
Equipment
tab
Select the Measurement / Display Device
Serial Number from the drop down list.
Select the Transducer Serial Number from
the drop down list.
Note: If using the Measurement Devices integral
transducer, leave Transducer Serial Number blank.
Check the current certificate numbers are correct
and that the uncertainty values are correct.
If not, correct them in the Measurement Device
and Transducer registers.