61
10. If adding an additional conformance the
Certificate Text
screen will appear.
Use the pull down option
to select
As found
,
As left
or enter your own text.
Select OK.
11. Select the
Equipment
tab to review the calibration
equipment used to perform the conformance test
Note: Check the current certificate numbers are correct and that the uncertainty values are correct.
If not, correct them in the Measurement Device and Transducer registers.
12. Select the
Readings
tab to review the conformance test readings
Press the ‘Create Certificate’ button.