CONTROL settings
the same port for transmission as for reception (i.e. as configured under the
System
port
). Default setting is 0. If
Local port
in CONTROL’s Ethernet connection
is configured different than the
Remote
port
, the computer’s port should match
the
Remote port
.
MAC Address
The
MAC address
for the laser’s Ethernet frames is shown in the bottom of the
Ethernet settings. The
MAC address
is not configurable but unique for every
laser.
Figure 47
Ethernet setting
Watchdog
The laser features a watchdog for monitoring of USB or Ethernet communication.
If CONTROL is disconnected from the chassis, the watchdog disables emission
after its timeout expires.
Enable Watchdog
To turn the watchdog on, select ON from the drop-down menu next to
Enable
watchdog
. To turn the watchdog off, select OFF.
Watchdog timeout
When CONTROL communications is disconnected, the
Watchdog
timer starts.
When it reaches the
Watchdog timeout
setting, emission is disabled. Set
Enable
Watchdog
to ON to enable the feature.
Содержание Koheras ADJUSTIK HP
Страница 1: ...Item 800 611 01 Koheras ADJUSTIK HP PRODUCT GUIDE Low Noise Single Frequency Laser System...
Страница 10: ...10 E Part numbers 113...
Страница 12: ...12...
Страница 16: ...16...
Страница 18: ...18...
Страница 20: ...20...
Страница 36: ...Chassis labels 36...
Страница 48: ...48...
Страница 56: ...Menu items 56...
Страница 68: ...Enabling and disabling emission 68...
Страница 84: ...CONTROL Control panel 84...
Страница 86: ...86...
Страница 90: ...Placing the laser on a table or shelf 90...
Страница 98: ...98...
Страница 102: ...102...
Страница 112: ...Installing CONTROL 112...
Страница 114: ...114 2...
Страница 115: ...Koheras ADJUSTIK Product Description Release 1 0 04 2021 W 10456...