CONTROL settings
76
“Wavelength modulation coupling” on page 42
information.
Note:
When using wavelength modulation with an X15 laser, and the Modulation
range parameter is set to
Narrow
, the modulation signal will be AC-coupled, and
the DC-component enters a software integrating function that tunes the wave
-
length up or down when the DC-voltage varies either positive or negative re
-
spectively. The speed of wavelength tuning is set with the Integrating update
interval.
Modulation
Range
For X15 lasers, you can use the Modulation
Range
drop-down menu to set the
modulation range to either narrow or wide.
When set to
Narrow
, modulation depth is limited but the X15’s low phase noise is
maintained. To increase the modulation depth, change the range setting to
Wide
.
When the modulation range is set to
Wide
, phase noise increases.
Note:
Refer to
“Narrow vs. Wide modulation range” on page 42
for further infor
-
mation on configuring the modulation range.
Signal output
You can output the internal generator signal (internal modulation) from the
wavelength modulation pins of the laser. To output the signal from the pins, set
the Modulation type to
Internal
and the
Signal
output
selector to
Enabled
. To
disable the output signal, set the
Signal
output
drop-down menu to
Disabled
.
When multiple lasers are used, one laser can be used as a master signal
generator with the other lasers configured as slaves.
1. For all the lasers, connect the
Wav
pins together and the
Wavelength-
pins together.
2. Designate one laser as the master by setting it to operate in
Internal
modulation mode with the
Signal
output
setting
Enabled
.
3. All other lasers will operate in slave mode by setting them to operate in
External
modulation mode.
Integrating update interval
You can set the update interval of the software integration function with this
setting. The setting is intended for use with X15 lasers when they are set to
DC
coupling and
Narrow
modulation range.
Internal generator
Frequency
When the modulation source is set to
Internal
, the frequency of the internal
generator signal can be adjusted to between 8 MHz (0.008 Hz) and 100 kHz
(100,000 Hz). To adjust the frequency, use either the slider or the direct input
field in the upper right corner of the slider.
Содержание Koheras ADJUSTIK HP
Страница 1: ...Item 800 611 01 Koheras ADJUSTIK HP PRODUCT GUIDE Low Noise Single Frequency Laser System...
Страница 10: ...10 E Part numbers 113...
Страница 12: ...12...
Страница 16: ...16...
Страница 18: ...18...
Страница 20: ...20...
Страница 36: ...Chassis labels 36...
Страница 48: ...48...
Страница 56: ...Menu items 56...
Страница 68: ...Enabling and disabling emission 68...
Страница 84: ...CONTROL Control panel 84...
Страница 86: ...86...
Страница 90: ...Placing the laser on a table or shelf 90...
Страница 98: ...98...
Страница 102: ...102...
Страница 112: ...Installing CONTROL 112...
Страница 114: ...114 2...
Страница 115: ...Koheras ADJUSTIK Product Description Release 1 0 04 2021 W 10456...