65
Enabling and disabling emission
Procedure 4
Turning on the laser using the front panel controls
Action
1 Connect to the laser using the procedures in
Communicating with the Laser on page 57
.
2
a.
If the interlock circuit is in an open state or
not terminated with the included External
bus defeater – the
External interlock
message is displayed.
To clear it, close all bus connections and
place a bus defeater on the External bus
connection of the bus chain’s last
connected device’s.
Note
if the laser is operating in standalone
mode, place the bus defeater on the
External bus connector.
b. Door interlock
message – the message is
displayed when the door interlock circuit is
open. Close the circuit to clear the
message.
Note
After closing all interlock circuits,
ensure to acknowledge the alarm using
Acknowledging interlock alarms on page 92
3 If the
Key switch off
message is displayed –
go to step 4.
4 Turn the key switch on the front panel to the
ON position.
External bus chain
or door interlock
circuit open
Front panel key
switch is in the
OFF position
ON
OFF
Содержание Koheras ADJUSTIK HP
Страница 1: ...Item 800 611 01 Koheras ADJUSTIK HP PRODUCT GUIDE Low Noise Single Frequency Laser System...
Страница 10: ...10 E Part numbers 113...
Страница 12: ...12...
Страница 16: ...16...
Страница 18: ...18...
Страница 20: ...20...
Страница 36: ...Chassis labels 36...
Страница 48: ...48...
Страница 56: ...Menu items 56...
Страница 68: ...Enabling and disabling emission 68...
Страница 84: ...CONTROL Control panel 84...
Страница 86: ...86...
Страница 90: ...Placing the laser on a table or shelf 90...
Страница 98: ...98...
Страница 102: ...102...
Страница 112: ...Installing CONTROL 112...
Страница 114: ...114 2...
Страница 115: ...Koheras ADJUSTIK Product Description Release 1 0 04 2021 W 10456...