Power On/Off
NOSEPIECE
C-HGFIE
USB
LV-AF
TE2-PS100W
DIA
TE2-PS100W
EPI
UEPI2A
100-240V
~
1.2A 50/60Hz
LV-FMA
LV-DICA
Rear of the
LV-ECON
Front of the
LV-ECON
Power switch
POWER indicator
CAUTION
When the power supply for the halogen lamp or the external light source is used with
this product, the power-on procedure is the following: Turn on the external device.
And then, turn on this product.
Particularly, it takes several minutes to start the external light source. Before turning n
the product, check that the external light source startup procedure has completed.
When the C-HGFIE is used for the light source and has completed the startup
procedure, the LAMP indicator changes from blinking to lighting. When the EXFO X-
Cite120 PC is used for the light source and has completed the startup procedure, the
backlight of the LCD changes from blinking to lighting.
Power ON/Off
The power switch of this product is located beside the AC inlet on the rear. To turn on the product,
push the power switch to the “
I
” side. To turn off the product, push the power switch to the “ ”
side. When the product is turned on, the POWER indicator lights green.
•
Initialization of the microscope system
When the product is turned on, a beep sounds, all indicators on the operation panel light up, and this
product starts communications with each electrical device. When the communication is established,
each electrical device is set to the predetermined conditions and the indicators on the operation
panel show the current condition. This procedure is called the “initialization” of the microscope
system.
When the initialization ends, a beep sounds again to inform that this product is ready for use. Under
the factory default condition, it takes about 15 to 20 seconds to initialize the microscope system.
The time for the initialization varies depending on the settings of the setup.
* If communications to a device fails in the initialization, the indicator on the operation panel turns
off to inform that the device is disabled.
Microscopy method setting at the initial condition
When the microscopy method at the initial condition has been set with LVSetup and the LV-ECON
is turned on, the microscope system starts with the predetermined microscopy method settings.
When the microscopy method at the initial condition has been disabled with LVSetup, the
microscope system starts with the same microscopy method settings used at the last time.