Selecting the Microscopy Method
When the LV-UEPI2A epi illuminator is connected with this product, one of the CUBE indicators
lights up to show the address of the filter cube turret in the optical path. When you wish to change
the microscopy method, push the desired button from the BF / DF / FL1 / FL2 buttons in the CUBE
area to rotate the filter cube turret.
Selecting the microscopy method
When one of the BF / DF / FL1 / FL2 buttons in the CUBE
area is pushed, the filter cube turret in the LV-UEPI2A
rotates to the position for the specified microscopy
method. The current microscopy method is indicated on
the indicator of the CUBE control and on the microscopy
method indicator of the LV-UEPI2A.
Control
BF
DF
FL1
FL2
Filter cube selection
Bright-field block
Dark-field block
Epi-fl filter cube (or PA cube) at the FL1
block
Epi-fl filter cube at the FL2 block
Interlock function for the microscopy method change
The interlock control is a function to change the electrically-driven devices of the microscope
system to the predetermined condition referring to the microscopy method and the objective when
the microscopy method or the objective is changed.
Selecting the microscopy method
BF
DF FL1 FL2
FL1
FL2
Microscopy method indicators
When the interlock control is enabled with LVSetup, the following devices are changed to the
predetermined conditions in connection with the objective and the microscopy method.
•
Aperture diaphragm size for the episcopic illumination
•
Episcopic illumination (on/off control and brightness control)
•
Diascopic illumination (on/off control and brightness control)
•
Vertical position of the motorized focusing module (parfocal correction)
•
Motorized DIC nosepiece (DIC prism insertion/removal, position A/B, fine adjustment position)
CAUTION
If you don’t perform the setup work, the microscope system does not work correctly.
Be sure to setup the microscope as follows.