![background image](http://html1.mh-extra.com/html/nikon/lv-econ/lv-econ_instructions-manual_1690699023.webp)
21
III. Device Connections
6
To Connect with the Light Source for the Diascopic Illumination
When a halogen lamp is used as a light source for the diascopic illumination, the lamp house (LV-
LH50PC) and the power supply (TE2-PS100W) must be connected. And then, the power supply
and this product must be connected with a control cable.
1
Check that the lamp house is correctly attached to a microscope base that can be equipped
with a diascopic illuminator.
2
Check that the power switch on the power supply is turned off (the “ ” side is pushed).
3
Connect the special lamp cable to the lamp cable of the lamp house and to the OUTPUT
connector on the power supply.
4
Connect the special control cable, which is 2.8 m in length, to the EXTERNAL connector on
the power supply and to the DIA TE2-PS100W connector on this product.
Make sure that the control cable is connected to the connector labeled DIA.
5
Move the EXTERNAL switch on the back of the power supply to the ON side.
MADE IN JAPAN
MODEL TE2-PS100W
ON
OFF
E
X
T
E
R
N
A
L
C-HGFIE
USB
LV-AF
TE2-PS100W
TE2-PS100W
UEPI2A
100-240V
~
1.2A 50/60Hz
LV-FMA
LV-DICA
NOSEPIECE
DIA
EPI
TE2-PS100W
LV-LH50PC
LV-ECON
Lamp cable
Control cable
* An optional control cable of 10 m in length is also available.
For details about installation method of the lamp house and replacement method of lamps,
refer to the instruction manual for the epi illuminator (LV-UEPI, LV-UEPI2, or LV-
UEPI2A).