Chapter 4 Setting Up the Microscope
4.2 Setup Procedure
1-87
4.2.3
Attaching the Laser Safety Kit
Refer to the instruction manual included with the microscope main body before attaching the laser safety kit.
Figure 4.2-25
1.
Attach the filter slider cover to the dia pillar
illuminator 100W.
Put the filter slider cover from above the dia
pillar illuminator 100W and secure it using the
screws and nuts provided with the laser safety
kit.
2.
Attach the protection plate for nosepiece
between the epi-fl filter turret and the nosepiece.
Notes on attaching the nosepiece --- When the epi-fl filter turret is used
The protection plate for nosepiece is a component designed for shielding the gap between
the nosepiece and the epi-fl filter turret to avoid accidental jamming of fingers inserted into
the gap or the laser radiation from the gap.
Two types of protection plate are prepared for the nosepiece being used. One for the TI-N6,
TI-ND6 and TI-ND6-E nosepieces is supplied with Ti-E and Ti-E/B, and the other for the
TI-ND6-PFS nosepiece is supplied with the nosepiece. Follow the procedure below to attach
each protection plate.
If the Perfect Focus System is not used
CAUTION
– CLASS 3B
LASER RADIA
TION WHEN OPEN
AVOID EXPOSURE T
O THE BEAM
VORSICHT
– LASERSTRAHLUNG
KLASSE 3
B,
WENN ABDECKUNG GEÖFFNET
NICHT DEM STRAHL A
USSETZEN
AVOID EXPOSU
RE
LASER LIGHT
IS EMITTED
FROM THIS A
PERTURE
N
AM
G
ET
N
Figure 4.2-26
1) Remove the two M3 hexagon socket head
bolts secured on top of the epi-fl filter turret
(TI-FLC, TI-FLC-E, or TI-FLC-E/HQ), using
the 2-mm hexagonal screwdriver appended
to the Ti-E and Ti-E/B.
2) Attach the protection plate (for nosepiece)
appended to the Ti-E or Ti-E/B to the top
portion of the epi-fl filter turret and secure
the plate using the two M3 hexagon socket
head bolts removed in step 1).
3) Attach the nosepiece (TI-N6, TI-ND6, or
TI-ND6-E) to the Ti-E or Ti-E/B and secure
it using the 4-mm hexagonal screwdriver
appended to the Ti-E and Ti-E/B.
Filter slider cover
Screws and nuts
(both side)
Dia pillar
illuminator
Protection plate
(for nosepiece)
TI-N6 (6 nosepiece),
TI-ND6 (DIC 6 nosepiece), or
TI-ND6-E (motorized DIC 6 nosepiece)
M5 Hexagon
socket head
bolts (x 2)
M3 Hexagon
socket head
bolts (x 2)
Содержание Eclipse Ti Series
Страница 2: ......
Страница 13: ...Part 1 For Setting Up TI TIRF TI TIRF E or TI PAU Series Setup Manual For Authorized Nikon Personnel ...
Страница 14: ......
Страница 256: ......
Страница 257: ...Part 2 For Setting Up TI TIRF PAU and LU4 B5 Series Setup Manual For Authorized Nikon Personnel ...
Страница 258: ......
Страница 260: ...Contents 2 2 8 2 Environmental Conditions 2 70 8 3 Safety Standards Compliance 2 71 ...