Chapter 6 Troubleshooting
6.2 Mechanical System
67
6.2
Mechanical System
Problem
Possible causes
Remedy
The image is not in focus
although the objective is
raised to the highest
position.
The stage has been attached
incorrectly.
Install the stage correctly.
The specimen is placed upside-down.
Set the specimen on the stage with the
cover glass facing up.
Focusing is not possible
with high-power
objectives.
The cover glass is too thick.
Use the specified type of cover glass
(thickness: 0.17 mm).
The specimen is placed upside-down.
Set the specimen on the stage with the
cover glass facing up.
The cover glass is too thick.
Use the specified type of cover glass
(thickness: 0.17 mm).
The objective strikes the
specimen when
changing from a
low-power objective to a
high-power objective.
The diopter adjustment has not been
performed.
Perform the diopter adjustment.
The interpupillary distance has not
been adjusted.
Adjust the interpupillary distance.
When viewing through
the binocular eyepiece,
the image does not
resolve into a single
image.
The diopter adjustment has not been
performed.
Perform the diopter adjustment.
The interpupillary distance has not
been adjusted.
Adjust the interpupillary distance.
The diopter adjustment has not been
performed.
Perform the diopter adjustment.
Eye strain develops
while viewing.
The brightness is not suitable.
Adjust the brightness with the brightness
control knob or ND filters.
6.3
Electrical System
Problem
Possible causes
Remedy
There is no power even
though the power switch
is on.
The power cord is not connected, or
is connected improperly.
Connect it properly.
The lamp does not light.
No lamp is attached, or the lamp has
burned out.
Replace the lamp with a specified one.
The lamp burns out
quickly.
The lamp type is incorrect.
Replace the lamp with a specified one.
The lamp has reached the end of its
operational life.
Replace the lamp with a new one.
The power cord is not connected
securely.
Connect it properly.
The lamp flickers; the
brightness is unstable.
The lamp has not been plugged into
its socket securely.
Install the lamp correctly.