Chapter 4 Assembly
59
7
Attaching the lamphouse for the epi illuminator
•
To prevent electrical shock and damage to the microscope, always turn off the
power switch (flip it to the “O” side) and unplug the power cord from the
outlet before attaching or detaching the lamphouse.
•
To prevent burn injury, allow the lamp and the lamphouse to cool down
sufficiently (for at least 30 minutes after the lamp is turned off), before
replacing the lamp.
•
Use the Nikon LV-LH50PC halogen lamphouse for the lamphouse.
(1)
Loosen the clamp screw on the top side of the
lamphouse connector by using the hexagonal
screwdriver.
(2)
Mount the lamphouse to the connection port on
the rear of the illuminator and press the
lamphouse as far as it goes.
(3)
Using the hexagonal screwdriver supplied with
the microscope, tighten the clamp screw on the
top of the connection port of the lamphouse to
secure it.
(4)
Connect the cable of the lamphouse with the
connector on the power supply device using an
extension cord.
CAUTION
Mount the lamphouse
lamphouse
Tighten the clamp csrew.
To power
supply
The power line connection for the power supply
device must be performed after attachment steps
for other devices. For details, refer to the
instruction manuals provided with the power
supply device.