Chapter 3 Individual Operations
3.16 Polarization Microscopy
46
3.16.2
Operation of the Analyzer
Attach/detach the analyzer
The intermediate tube has the analyzer slider. The
analyzer can be placed into the optical path with
the operation of the slider.
To place the analyzer into the optical path, pull out
the slider. To remove the analyzer from the optical
path, push in the slider.
The analyzer is designed to be inserted from the
right side in normal use, but it can be inserted from
the left. In the later case, its scale displays the
opposite way. Be careful.
Slider Analyzer
Pulled out
IN
Push in
OUT
Rotate the analyzer
The analyzer slider has a rotating dial. The
orientation of the analyzer can be rotated with it.
To adjust the analyzer, loosen the analyzer rotation
clamp screw and rotate the rotating dial.
The angle of rotation can be read from 0 to 180
degrees in steps of 0.1 degrees with the two
vernier scales.
0
10
OUT
IN
0.
120
110
100
130
30
20
10
40
80
70
60
50
90
170
160
150
140
0
A
The intermediate tube also has a de-polarizer. So, you can use the
photomicrography devices independently of the orientation of the polarizer.
Analyzer rotation
clamp screw
Analyzer
slider
Vernier
Analyzer
rotating dial