16
Microscopy
2
.
2.1
Diascopic Bright-field Microscopy
This section describes the diascopic bright-field microscopy using the built-in lamp of the
microscope.
1
Turn on the power.
When the microscope has power applied to it, the
power indicator on the right side of the microscope
is lit.
⇒
P.29
2
When the episcopic illuminator is
attached, turn off the episcopic
illuminator. And then, pull the
illumination selection lever to set
the dark-field position for it.
⇒
P.51
3
Rotate the brightness control knob
to adjust the brightness.
Or, push the preset switch to apply the voltage that
wil make good color reproducibility.
⇒
P.30
4
Push in the analyzer slider on the
polarizing intermadiate tube to
remove the analyzer from the
optical path.
⇒
P.46
Remove the
analyzer from
the optical path.
Push the power switch to
the “I” position.
Adjust the brightness, or
push the preset switch.
Set to the DF position.