18
CHECK AND ADJUSTMENT PROCEDURES
:
Size Check and Adjustment
3.3
Size Check and Adjustment
3.3.1
Checking the size
1
Load the circle 45 internal data.
For the circle 45 internal data, see “3.2.2
Loading the shape for adjustment” (page
17).
Process a lens with the settings as Two
Point, Flat, None (Polish), None (S.F.B.),
and Size 0.00.
2
Measure the outer diameter of the processed
lens with a caliper and check the size.
The outer diameter of the flat-edged lens is within
ø45.00mm ±0.05.
3.3.2
Adjusting the size
1
Display the menu screen in service mode.
See “3.2.1 Entering service mode” (page 16).
Outer diameter
Within ø45.00 mm
±0.05
• Calibrate surface measurement (LMU) once when the size is outside the range.
See “3.4 Surface Measurement (LMU) Calibration” (page 20).
• After the surface measurement (LMU) is calibrated, load the circle 45 internal data and
process a lens again. When the size is outside the range, adjust the size in the following
procedure.
Содержание ME 1200
Страница 6: ...IV ...
Страница 10: ...2 REQUIREMENTS Requirements ...