MAINTENANCE
:
Replacing Fuses
3.2
Replacing Fuses
If the system is not started even though the power switch is turned on, the fuses may be blown.
Replace the fuses with spare ones.
1
Turn the power off (
).
2
Remove the power cord from the electrical outlet.
3
Unplug the power cord from the inlet.
4
Remove the fuse holder under the inlet.
Pull out the fuse holder while pushing the
lever.
5
Take out the fuses and put in new ones.
Fuse rating: T 1 A 250 V
(AC 100 - 120 V/230 V)
6
Put the fuse holder back into its original posi-
tion.
CAUTION
• Be sure to turn off the power and disconnect the power cord from the outlet before
replacing the fuses.
Otherwise, electric shock may result.
• Only the specified fuses (T 1 A 250 V) must be used.
Any fuses other than the specified ones may cause fire.
Lever
Fuse holder
Fuse holder
Fuse