4.3 Basic Functions
BP4610
4-16
<3> Press the and
keys to move the cursor to the digit to be changed and then turn the
[MODIFY] dial to set the desired value. When the output is on, the value that is set is
immediately reflected to the output.
<4> Once all the digits have been set, press the
CANCEL
or
ENTER
key to end the setting procedure.
4.3.4 Setting of superimposed AC voltage and superimposed AC
current
To superimpose an AC voltage on the DC output voltage of the BP4610, or to output only an AC
voltage (CV-INT or CV-ADD mode), set the superimposed AC voltage via the operation panel.
•
Setting range:
0 Vp-p to 120 Vp-p
•
Resolution:
0.1 Vp-p
To superimpose an AC current on the DC current of the BP4610, or to output only an AC current (CC-
INT or CC-ADD mode), set a superimposed AC current via the operation panel.
•
Setting range:
0 Ap-p to 30 Ap-p
•
Resolution:
0.01 Ap-p
Operation procedure
<1> Press the
and
keys to move the cursor to the
icon (when superimposed AC
voltage is set), or the
icon (when superimposed AC current is set).
ENTER
CANCEL
Содержание BP4610
Страница 1: ...NF Corporation BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual nbn Austria GmbH...
Страница 2: ......
Страница 3: ...DA00010493 001 BIPOLAR DC POWER SUPPLY BP4610 Instruction Manual...
Страница 4: ......
Страница 16: ......
Страница 17: ...BP4610 1 OVERVIEW 1 1 General 1 1 1 2 Features 1 1 1 3 Principle of Operation 1 3...
Страница 18: ......
Страница 24: ......
Страница 38: ......
Страница 40: ......
Страница 48: ......
Страница 134: ......
Страница 162: ......
Страница 164: ......
Страница 167: ...6 1 Error Messages BP4610 6 3 Figure 6 1 Protective Function Related Error Display Examples...
Страница 174: ......
Страница 190: ......
Страница 207: ...BP4610 9 INITIAL SETTING LIST 9 1 Initial Setting List 9 1...
Страница 208: ......
Страница 212: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......