68
EVENTEN
Device Event Enable Register
Syntax:
EVENTEN <mask>
Parameters:
<mask> is an integer in the range 0 to 255 inclusive. The value of this number
is written into the Device Event Enable register.
Function:
The Device Event Enable register is AND’ed with the Device Event register. If
any bit is set in the result of this AND’ing operation then Device Error bit (bit 3)
in the Standard Event Status register will be set.
The Device Event Enable register is an 8 bit, bit mapped register. Any bit set
to 1 in the Device Event Enable register allows the corresponding bit in the
Device Event register to set the Device Error bit (bit 3) in the Standard Event
Status register. Any bit set to 0 disables the corresponding bit in the Device
Event register from setting the Device Error bit.
Over Range CH A
Saturated CH A
Data Error CH A
(Not Used)
Over Range CH B
Saturated CH B
Data Error CH B
(Not Used)
0
1
2
3
4
5
6
7
Device Event Register
Related Commands:
EVENT?, EVENTEN?, *ESE, *ESE?, *ESR?
EVENTEN?
Device Event Enable Register Query
Syntax:
EVENTEN?
Parameters:
None
Function:
This query returns the contents of the Device Event Enable register. See the
EVENTEN command for a description of the Device Event Enable register.
Returns:
<mask>
<mask> is of type <number> and represents an unsigned decimal integer in the
range from 0 to 255.
Related Commands:
EVENT?, EVENTEN, *ESE, *ESE?, *ESR?
EXT
External Trigger Enable
Syntax:
EXT <enable>
Parameters:
The parameter <state> is either 0 or 1. If <state> is 0 then the external trigger
is disabled. If <state> is 1 then the external trigger is enabled.
Function:
This command enables or disables the external trigger input on the back panel.
Related Commands:
EXT?, EXTEDGE, EXTEDGE?